General Description
Multichannel, interpretive electrocardiograph, including printer.

Technical Specifications :-
ECG analysis and full interpretation (rhythm and events), real-time and manual interpretation.
Three (3) or more recording channels.

Signal Acquiition :-
Simultaneous 12 lead acquisition ( aVR, aVL, aVF, I, II, III and V1 - V6) derived from 10 electrodes (RA, LA, RL, LL, V1 - V6).
Arrythmia and ST elevation detection.
Minimum gain/sensitivity settings include 2.5, 5, 10 and 20 mm/mV.
Lead switching, either manual, automatic or both (manual and auto options).
Adjustable trace speeds include 5, 10 (and/or 12.5), 25 and 50 mm/s.
Minimum HR range 30 – 300 bpm with rhythm analysis.
Minimum guaranteed diagnostic frequency response of 0.05 – 150 Hz.
Common Mode Rejection Ratio (CMRR) at 60 Hz > 105 dB or better.
Calibration signal of 1 mV, manual and/or automatic.
Selectable/adjustable filters for baseline drift, muscle artefacts, mains power.
Pacemaker detection.
Accuracy of input signal reproduction ± 5 % or ± 40 μV, whichever is greater.
Input impedance > 50 MΩ.
Internal noise level < 12.5 µV peak-to-peak.
Automatic baseline centring.
Defibrillation fluctuation/overload protection.
Baseline recovery < 5 s after defibrillation.
AC fluctuation protection.

Recorder and Printer :-
Minimum of 3 recording channels.
Recorder display includes date/time, patient data and heart rate and basic settings.
Capable of displaying one group of at least three channels simultaneously.
Recorder waveform display includes lead marker and timing marker.
Integrated/built-in printer.
Capable of printing user selected number of channels.
Capable of printing one group of at least three channels simultaneously.
Paper speeds include 5, 10 (or 12.5), 25 and 50 mm/s.
Compatible with Z-fold paper and optionally with roll paper also (indicate compatibility).

Data Input/Output, Storage and Alarms
Integrated alpha-numeric keyboard.
Patient data input fields include name, age, height and weight, gender.
Backlit Liquid Crystal Display (LCD) display screen, minimum of 12.7 cm (5 inches) diagonally.
Capable of internally storing a minimum of 10 waveforms for later retrieval, printing and/or transmission.
Expandable storage for additional waveforms if required, via USB.
Capable of exporting waveform data and reports, via USB or LAN.
Visual alarms for patient connection (lead faults), heart rate, printer and paper errors, and system errors.
Automatic self-test at start up.

Power Provisions :-
Built-in rechargeable lithium-ion battery.
Minimum battery operating time is 100 ECG exams or 4 hours of continuous recording.
Automatic switch to battery in case of power failure, automatic recharge on connection to mains.
Maximum battery charging time to full charge is 8 hours.
Power requirements : 100 - 240 Volts - 50/60 Hz.
Designed for frequent and easy dismount and disinfection with hospital-grade products.

Intended Use :-
Electrocardiographs detect the electrical signals associated with cardiac activity and produce an ECG, a graphic record of the voltage versus time. 
They are used to diagnose and assist in treating some types of heart disease and arrhythmias. 
Electrocardiographs differ from ECG monitors, which provide a continuous real-time display of the ECG waveform and other physiologic measurements on a screen and warn of potentially life-threatening or precursor rhythms. 
Electrocardiographs are used to diagnose cardiac abnormalities, determine a patient's response to drug therapy, and reveal trends or changes in heart function. 
Multichannel electrocardiographs record and display signals from two or more leads simultaneously.


   